GE CT systems at a basic service level. Note: Completion of this course or the Challenge Exam is required to attend a Full Service CT course. The systems taught in the course include: LightSpeed TM 4/8/16, LightSpeed 4/8/16 (HP60), LightSpeed 16 Pro, LightSpeed RT, LightSpeed VCT TM, BrightSpeed, HiSpeedTM NX/I. Course competencies. The 64-slice LightSpeed VCT XT helps patient breath-holds to. The GE LightSpeed VCT XT brings exceptional CT imaging power to the Discovery PET/CT 690.

  1. Ge Lightspeed Vct 64 User Manual Pdf
  2. Ge Lightspeed Vct 64 User Manual 1
  3. Ge Lightspeed Vct 64 User Manual Download
  4. Ge Lightspeed Vct 64 Manual
  5. Ge Ct Lightspeed Technical Manual
  • This program instructs CT Technologists in the theory and operation of the GE LightSpeed VCT CT system. Instruction is delivered by a GE CT Applications Specialist at the customer facility and focuses on intermediate system operation and patient procedures.
  • GE HEALTHCARE DIRECTION DOC0860282, REV 3 BRIGHTSPEED AND LIGHTSPEED DICOM CONFORMANCE STATEMENT 4 of 172 REVISON HISTORY REV. DATE REASON FOR CHANGE 1 2 3 7/26/2011 New document number and product/sw version based on DOC0793025 Rev1.
  • GE HEALTHCARE DOC01200461, REV 3 LIGHTSPEED VCT 7.2 AND LIGHTSPEED VCT SELECT WITH 64 CHANNEL UPGRADE DICOM CONFORMANCE STATEMENT 4 of 185 REVISON HISTORY REV. DATE REASON FOR CHANGE 1 17/Aug./2012 Used DOC1053442 Rev 2 as baseline to create LightSpeed VCT 7.2 and LightSpeed VCT Select with 64 channel upgrade Dicom Conformance.

GE Lightspeed VCT no load software

July 16, 2014 04:20

When you turn the computer starts the operating system without problems, but when running 'startup' I get a window with the following message:
[cupMonitor] GEOCommServer: created listen channel: inet_stream 192.9.101.30 51947|@tm
[cupMonitor] CupMonitorService : com.ge.med.fct.coreplatform.startup.CupMonitorEvent has been advertised by 1;serv=inet_stream 192.9.101.30 51947|@tm
chmod: changing permissions of `/tmp/.genameservice/': Operation not permitted
chmod: changing permissions of `/tmp/.genameservice/GEOCOMM': Operation not permitted
chmod: changing permissions of `/tmp/.genameservice/GEOCOMM/cupMonitor': Operation not permitted
----------------------------------------------
I need to know why this happens.

Ge Lightspeed Vct 64 User Manual

re: GE Lightspeed VCT no load software

July 17, 2014 01:24

Have you try LFC

re: GE Lightspeed VCT no load software

July 17, 2014 01:25

Hi
The erro message is :
----------------------------------------------
CUP ERROR in ../../source/cupMakeCore.c line 200 pid 528
generating traceback file traceback.cupMonitor.201416Jul.120517
----------------------------------------------

re: re: GE Lightspeed VCT no load software

July 17, 2014 08:40

That could mean anything, your clue is that it fails at start up, if you don't want to do a LFC try a restoring system state first

re: re: re: GE Lightspeed VCT no load software

July 17, 2014 05:58

what is LFC?

re: re: re: re: GE Lightspeed VCT no load software

July 17, 2014 09:19

Load from cold , reload software on the computers

re: re: re: re: GE Lightspeed VCT no load software

July 17, 2014 09:19

Load from cold, reload software in the computers

Lightspeed

re: re: re: GE Lightspeed VCT no load software

July 17, 2014 05:59

the gesys_vct_cdd.log show this error:
SR 204
1405339783 0 1 Mon Jul 14 12:09:43 2014 0 7
vct_cdd Svc_Notepad
Notepad.c 44
This message was added by the OPERATOR to report on a problem:
startMon failed to mount oc:/usr/g on DARC
EN 204
SR 389
1405341058 0 1 Mon Jul 14 12:30:58 2014 0 7
vct_cdd Svc_Notepad
Notepad.c 44
This message was added by the OPERATOR to report on a problem:
PRODUCT CONFIGURATION|Product: LightSpeed VCT|sw_version: 12HW14.6|sw_hw_string: 12HW14.6_SP1-1-1.V40_H_H64_G_GTL|host_id: 46211168|tube_type: 16|gantry_type: GANTRY_VCT_40|No. of processors: 4|Processor speed: 3200|memory: 4096
EN 389

Ge Lightspeed Vct 64 User Manual Pdf

re: re: re: re: GE Lightspeed VCT no load software

July 18, 2014 10:04

It seems that your DARC is not working properly.
Try this Linux commands to see if you can communicate with it :
- ping darc
- rsh darc
If these commands are OK, your issue can be a bad DARC disk drive

Log inor Register

to rate and post a comment

re: re: re: GE Lightspeed VCT no load software

March 01, 2016 09:27

hi! i have trouble with light speed 8 slice with Goc 4 console.
i need reload the software do you know how to do?
i put the DVD disc on the cpu i start the console and it put in boot mode and gave me some options to choice, one of them is recovery system , i can do this?
the problema is i have to replace the darc, but it not had the software installed.

Log inor Register

to rate and post a comment

re: re: re: re: GE Lightspeed VCT no load software

March 02, 2016 09:12

Hi Miguel you need to load only the software for your darc, no LFC if you only Replace the darc.
sysmedi22@gmail.com I am in México.

re: re: re: re: GE Lightspeed VCT no load software

Lightspeed

March 02, 2016 09:24

LFC should solve all your problems. Yes, load all the 3 disks and then the fourth is the applications disk. Do you know the passwords? I could send them to you if you need them. #bigbuy & 4$apps

Ge Lightspeed Vct 64 User Manual 1

re: GE Lightspeed VCT no load software

Ge Lightspeed Vct 64 User Manual Download

July 17, 2014 05:58

Dear
What is version of your console: GOC5, 5.5 or 6?
First: You check Ethernet card
Second: reconfig it
Third: LFW it
If three step above is not solved your problem, you can do LFC
Good Luck

Log inor Register

Ge Lightspeed Vct 64 Manual

to rate and post a comment

re: re: GE Lightspeed VCT no load software

July 18, 2014 10:03

Ge Ct Lightspeed Technical Manual

Hello, I agree with Tuy Do Duy,
If you need to know the version of your console, open a unix shell and type :
grep CONSOLETYPE /usr/g/config/INFO
If you have a GOC 5 or GOC 5.5, check if the host communicates with the DARC (use the following command : ping DARC) or look if the DARC has booted.
Good luck